Catalog Number 0048-0024-JP
Medical Device Problem Codes Restricted Flow rate (1248); Malposition of Device (2616); Patient Device Interaction Problem (4001)
Health Effect - Clinical Codes Low Blood Pressure/ Hypotension (1914); Mitral Valve Insufficiency/ Regurgitation (4451)
Date of Event 12/01/2024
Type of Reportable Event Serious Injury
Event or Problem Description
Clinical narrative: an impella cp was inserted via the femoral artery, of undocumented side, to support the 75-year-old male patient admitted with known coronary artery disease and plan for the off-pump coronary artery bypass (opcab) surgery.The patient did present per documentation in society for cardiovascular angiography and interventions (scai) stage c shock at the time of pump insertion.The patient had a medical history inclusive of hypertension, dyslipidemia, chronic kidney disease, and heart disease.The impella cp supported the patient for 4 days in (b)(6) of 2024, and was successfully weaned and explanted.The team in japan published upon the support.The publication was discovered by a social media post in july 2026.The publication noted the impella cp was placed pre-operatively and supported for the opcab.While in the operating suite the pump became malpositioned deeper within the left ventricle.The positioning of the pump allowed for the mitral valve leaflet to be interacted with and the team noted new mitral valve regurgitation.After 90 minutes the patient condition deteriorated and the patient was noted to be hypotensive.The team made the decision to escalate care and extra corporeal membrane oxygenation (ecmo) was added.Once ecmo was added the impella cp acted as a ventricle ventilation for the primary support device, ecmo.The heart bypass surgery continued.At the conclusion of the opcab, the ecmo was explanted.The team further evaluated the impella pump position and the team saw there was no need for mitral valve surgery and repair.The pump was explanted after 4 days of support and on day 13 the patient was discharged from the intensive care unit.The mild jet observed from the mitral valve was still present, but did not need surgical intervention.The mitral valve regurgitation was not known to the abiomed team in japan in 2024 when the harm took place.Per the publication and medical team in japan, the use of the impella is useful to surgical patients however the change in pump position caused the impella to interfere with the posterior mitral valve leaflet and led to severe mitral regurgitation.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by abiomed inc, or its employees that the report constitutes an admission that the product, abiomed inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Impella
Common Device Name
Temporary Non-Roller Type Left Heart Support Blood Pump
Manufacturer (Section D)
Abiomed, Inc. - 1220648
22 cherry hill drive
danvers MA 01923
Manufacturer (Section G)
Danvers
Manufacturer Contact
derek sammarco
22 cherry hill drive
danvers, MA 01923
+1(781)493-4967
MDR Report Key50054905
Report Number1220648-2026-10960
Device Sequence Number15825397
Product Code OZD
UDI-Device Identifier00813502011371
UDI-Public(01)00813502011371(17)260430(10)2025511538
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P140003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Literature,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2026
Device Catalogue Number0048-0024-JP
Device Lot Number2025511538
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/20/2026
Supplement Date Received by Manufacturer07/28/2026
Initial Report FDA Received Date07/28/2026
Supplement Report FDA Received Date07/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/28/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ECMO
Outcome Attributed to Adverse Event Required Intervention;
Patient Age75 YR
Patient SexMale
Patient Weight125 LB
Patient RaceAsian
-
-