| Catalog Number |
500-03-52D |
| Medical Device Problem Code |
Loss of Osseointegration (2408)
|
| Health Effect - Clinical Code |
Inadequate Osseointegration (2646)
|
| Date of Event |
11/24/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The following information was provided: it was reported through the filing of a lawsuit that allegedly the patient was implanted with an lfit anatomic cocr v40 femoral head on her left hip on or about (b)(6)2016 and was revised on (b)(6)2021.It is further alleged that she suffered injuries as a result of implantation and explantation of the device at issue, and device recall.The revision operative report states that the acetabular component was grossly loose and easily removed.It was revised to a larger shell.Upon removal of the femoral head corrosion with black material was noted on the trunnion.Additional info - med review states 'the revision operative report states that the acetabular component was grossly loose and easily removed.It was revised to a larger shell.Upon removal of the femoral head corrosion with black material was noted on the trunnion.'.
|
| |
|
Additional Manufacturer Narrative
|
|
Review of the device history records indicate that devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.
|
| |
|