Aspen Surgical Products, Inc. RapidClean® Rongeur, 40° Up, Ceramic Coated, Convertible, 3
|
Back to Search Results |
|
| Model Number |
59-3200-BD |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The sample was returned for evaluation.Visual inspection showed that the gold screw is loose.The customer indicated that they are unsure whether the ssu adjusted the screw prior to sterilization.During manufacturing, loctite is applied to the screw threads, and the screw is also secured with a welding spot to achieve proper jaw alignment by locking the instrument in place so that it does not become unassembled as shown in the customer image.The root cause is user error, for removing the screw during ssu, which broke the loctite and spot welds resulting in the gold screw becoming loose after ightening.If any additional relevant information is identified it will be submitted in a supplemental report.
|
| |
|