Exactech, Inc. THREE PEG PATELLA 29MM; Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal
|
Back to Search Results |
|
| Catalog Number |
200-02-29 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Pain (1994); Scar Tissue (2060); Synovitis (2094); Discomfort (2330); Osteolysis (2377); Ambulation or Postural Difficulties (2544); Joint Laxity (4526); Swelling/ Edema (4577)
|
| Date of Event |
01/23/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
According to the information provided, the patient underwent a left total knee replacement surgery.Approximately 12 years and 11 months later, the patient had revision surgery on the left knee.All components were removed.Per the operative report, the patient presented with significant mechanical loosening and instability of their left primary total knee prosthesis.This was refractory to comprehensive nonoperative management due to extensive polyethylene wear and osteolysis.Because the patient has filed a long form, it appears that they are alleging prosthesis wear, significant pain and discomfort; gait impairment; poor balance; difficulty walking; component part loosening; soft tissue damage; bone loss; and other injuries presently undiagnosed.
|
| |
|
Search Alerts/Recalls
|
|
|