MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION Wolverine Coronary Cutting Balloon; catheter, percutaneous transluminal coronary angioplasty (ptca), cutting/scoring

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION Wolverine Coronary Cutting Balloon; catheter, percutaneous transluminal coronary angioplasty (ptca), cutting/scoring Back to Search Results
Model Number H74939401153000
Medical Device Problem Codes Break (1069); Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/29/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the balloon wires broke apart.A two 15mmx3.00mm wolverine coronary cutting balloon were selected for treatment.During the procedure it was noted that the balloon wires broke apart.Removal of the wire percutaneously were unsuccessful.An incision had to be made to remove the broken wires.There was no patient complications reported.It was further reported that this case was reported in medwatch# mw5190435.
 
Event or Problem Description
It was reported that the balloon wires broke apart.A two 15mmx3.00mm wolverine coronary cutting balloon were selected for treatment.During the procedure it was noted that the balloon wires broke apart.Removal of the wire percutaneously were unsuccessful.An incision had to be made to remove the broken wires.There were no patient complications reported.
 
Additional Manufacturer Narrative
With all the available information, boston scientific concludes: device analysis findings: the device was not returned for analysis; therefore, a technical analysis could not be performed.Device history record (dhr) review: review confirmed that the device met all material, assembly, and performance specifications prior to distribution.Labeling review: review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.Risk review: a risk review was performed and confirmed that the event was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Investigation conclusion: this investigation is assigned a most probable investigation conclusion code of cause not established.During the procedure it was noted that the two balloon wires broke apart and an incision had to be made to remove both wires.The device was not returned for analysis.Based on the limited information provided a clear conclusion of the reported event cannot be determined.
 
Additional Manufacturer Narrative
B3 date of event was estimated based on aware date and it being reported that the event took place the week prior.Good faith effort attempts were made to try and retrieve additional details regarding the reported event, but further information was unable to be obtained.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Wolverine Coronary Cutting Balloon
Common Device Name
catheter, percutaneous transluminal coronary angioplasty (ptca), cutting/scoring
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
ballybrit business park
galway
EI  
Manufacturer Contact
mark mischke
4100 hamline ave n
arden hills, MN 55112
+1(612)875-7234
MDR Report Key50059135
Report Number2124215-2026-40385
Device Sequence Number10730574
Product Code NWX
UDI-Device Identifier08714729888352
UDI-Public(01)08714729888352(17)280420(10)39145275
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P950020
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberH74939401153000
Device Catalogue NumberH74939401153000
Device Lot Number0039145275
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/06/2026
Supplement Date Received by Manufacturer08/13/2026
Initial Report FDA Received Date07/28/2026
Supplement Report FDA Received Date09/09/2026
Date Device Manufactured04/21/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-