| Model Number |
H74939401153000 |
| Medical Device Problem Codes |
Break (1069); Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/29/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the balloon wires broke apart.A two 15mmx3.00mm wolverine coronary cutting balloon were selected for treatment.During the procedure it was noted that the balloon wires broke apart.Removal of the wire percutaneously were unsuccessful.An incision had to be made to remove the broken wires.There was no patient complications reported.It was further reported that this case was reported in medwatch# mw5190435.
|
| |
|
Event or Problem Description
|
|
It was reported that the balloon wires broke apart.A two 15mmx3.00mm wolverine coronary cutting balloon were selected for treatment.During the procedure it was noted that the balloon wires broke apart.Removal of the wire percutaneously were unsuccessful.An incision had to be made to remove the broken wires.There were no patient complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
With all the available information, boston scientific concludes:
device analysis findings:
the device was not returned for analysis; therefore, a technical analysis could not be performed.Device history record (dhr) review:
review confirmed that the device met all material, assembly, and performance specifications prior to distribution.Labeling review:
review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.Risk review:
a risk review was performed and confirmed that the event was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Investigation conclusion:
this investigation is assigned a most probable investigation conclusion code of cause not established.During the procedure it was noted that the two balloon wires broke apart and an incision had to be made to remove both wires.The device was not returned for analysis.Based on the limited information provided a clear conclusion of the reported event cannot be determined.
|
| |
|
Additional Manufacturer Narrative
|
|
B3 date of event was estimated based on aware date and it being reported that the event took place the week prior.Good faith effort attempts were made to try and retrieve additional details regarding the reported event, but further information was unable to be obtained.
|
| |
|