| Model Number |
H74939401103000 |
| Medical Device Problem Codes |
Nonstandard Device (1420); Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/06/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the balloon burst.A 10mmx3.00mm wolverine coronary cutting balloon was selected for use.During the procedure, at first inflation, it was noted that the balloon burst.The device was removed without any problem using the normal method and the procedure was completed with another of the same device.There were no patient complications.
|
| |
|
Additional Manufacturer Narrative
|
|
With all the available information, boston scientific concludes:
device analysis findings:
the device was returned to the complaint investigation site for analysis.Visual/tactile inspection identified kinks in the hypotube shaft.Microscopic analysis observed distal tip damage (flared).Microscopic analysis also noted that the balloon appeared to have been subjected to positive pressure and blood was present in the balloon.The device was attached to an encore inflation device, and an attempt was then made to inflate the balloon however, pressure was unable to maintained as inflation liquid was observed leaking from a pinhole in the balloon material just proximal to the proximal markerband which was observed microscopically.Device history record (dhr) review:
it was confirmed that this device met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Labeling review:
a review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.Risk review:
a risk review was performed, and it confirmed that the event was defined in the risk documentation and is recorded accordingly in the product record review (prr) table.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Investigation conclusion:
this investigation is assigned a most probable investigation conclusion code of adverse event related to patient condition as it is most likely that anatomical factors (calcified and moderately tortuous vessel) were met which resulted in the reported event.
|
| |
|