BOSTON SCIENTIFIC NEUROMODULATION CORPORATION Infinion¿ CX; stimulator, spinal-cord, totally implanted for pain relief
|
Back to Search Results |
|
| Model Number |
SC-2317-50 |
| Medical Device Problem Codes |
High impedance (1291); Unexpected Therapeutic Results (1631); Charging Problem (2892)
|
| Health Effect - Clinical Code |
Pain (1994)
|
| Date of Event |
01/22/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient experienced a non-device related fall, after which the spinal cord stimulator (scs) leads exhibited new impedances, which resulted in inadequate pain relief.It was also noted that the patient experienced difficulty charging the implantable pulse generator (ipg), including a need for frequent charging.As a result, the patient underwent a procedure wherein the scs leads, ipg, and clik anchor were replaced.The patient was doing well postoperatively, and the explanted device components were not returned due to facility policy.
|
| |
|
Additional Manufacturer Narrative
|
|
Block b3: exact date unknown, event occurred two weeks prior to the date the manufacturer became aware of the event.Block d.2b; additional applicable product code: qrb.Additional suspect medical device components involved in the event:
model number/catalog number: sc-1232.Serial number: (b)(6).Batch/lot number: (b)(6).Model/catalog description: wavewriter alpha 32 ipg kit.Unique identifier (udi) #: (b)(4).Model number/catalog number: sc-2317-50.Serial number: (b)(6).Batch/lot number: (b)(6).Model/catalog description: nfinion cx lead kit, 50cm.Unique identifier (udi) #: (b)(4).Model number/catalog number: sc-4318.Serial number: na.Batch/lot number: 34955918.Model/catalog description: clik x anchor sterile kit.Unique identifier (udi) #: (b)(4).Sc-2317-50 sn: (b)(6).Investigation results
the spinal cord stimulator (scs) leads were not returned for analysis; therefore, a technical analysis could not be performed.Device history record
it was confirmed these devices met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Media review
review of the provided image did not contribute any relevant information to support or further evaluate the reported issue.Investigation conclusion
as the devices were not available for further evaluation and insufficient information exists to determine the underlying cause of the observed high impedances or their relationship to the reported fall, this investigation is unable to establish a definitive root cause.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.
|
| |
|