It was reported that the spinal cord stimulation (scs) patient presented with inadequate stimulation, described as insufficient pain relief in the targeted area.Evaluation determined that lead migration had occurred, resulting in the leads overlapping each other and contributing to the loss of coverage.Reprogramming was attempted several times with limited improvement.The patient underwent a revision procedure during which the existing leads were replaced with a paddle lead.During the same procedure, the implantable pulse generator (ipg) was replaced because the newly implanted paddle required a four port ipg.The explanted products were not released by the facility and were not returned for analysis.The patient is reported to be doing well postoperatively.
|
|
Block b3: exact date unknown, event occurred in december 2025.Additional suspect medical device components involved in the event:
device#: 2,
model number/catalog number: sc-1416,
serial number: (b)(6),
batch/lot number: 227563,
model/catalog description: wavewriter alpha prime 16 ipg kit,
unique identifier (udi)#: (b)(4),
device#: 3,
model number/catalog number: sc-2352-70,
serial number: (b)(6),
batch/lot number: 7083121,
model/catalog description: linear 3-4 lead 70 cm,
unique identifier (udi)#:(b)(4),
device#: 4,
model number/catalog number: sc-4318,
serial number: not applicable,
batch/lot number: 35501093,
model/catalog description: clik x anchor sterile kit,
unique identifier (udi)#: (b)(4).
|