It was reported that, following the implant of the transcatheter aortic valve, paravalvular leak (pvl) of unknown severity occurred.A post-implant balloon aortic valvuloplasty (bav) was performed while the patient was rapidly paced.Subsequently, the valve dislodged.Prior to valve dislodgement, the implant depth on the non-coronary cusp (ncc) and left coronary cusp (lcc) was 4 millimeter's (mm).After valve dislodgement, the implant depth on the ncc and lcc was -2 mm.The original valve was left in the sinus, and a second valve was implanted valve-in-valve at an implant depth of 6 mm.It was noted that mild pvl was observed post deployment, but a post-implant bav was not performed.Per the physician, the cause of the dislodgement was due to the post-implant bav and the depth of implant.
|
|
D10.Continuation - concomitant medical devices in use during the event include: medtronic product id: d-evolutfx-2329; product serial/lot number: (b)(6); product type: 0195-heart valves; implant date: not-imp lantable; explant date: na.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|