Apoc incident: 1010686
the investigation was completed on (b)(6) 2026.A review of the device history record confirmed the cartridge lot met finished goods release criteria.Although retained cartridge testing met the acceptance criteria outlined in appendix 1 of q04.01.003 rev.Ap (product complaint level 2 and level 3 investigation procedure), hs-tni cartridge lot c26140a is associated with a related previously identified deficiency for reduced interference mitigation robustness, which is captured in quality record 1130236.
|