Additional information: d9, g3, h3, h6.Correction: b5.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The involved device was not returned.Visual inspection noted that the returned product noted one suture and one needle.The needle was returned disengaged from the suture.The foil pouch was inspected and no signs of damage or abnormalities were identified.Upon microscopic inspection, instrument marks were noted near the swaged end of the needle.Suture material was observed to be present in the swage, indicating the suture broke off at the swage.Functional testing found that the opened complaint device was precluded as the needle was returned disengaged.It was reported that the needle detachment from the suture.The reported issue was confirmed.This issue can occurs if the needle is grasped near the swaged end or the tip and could cause bends, breaks, or damage the connection between the needle and suture.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: in handling this or any other suture material, care should be taken to avoid damage from handling.Avoid crushing or crimping damage due to application of surgical instruments, such as forceps or needle holders.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|