AngioDynamics, Inc. Smart Port CT; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
|
Back to Search Results |
|
| Model Number |
H787CT80STPD0 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Bacterial Infection (1735); Bacteremia (4846)
|
| Date of Event |
11/19/2015
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
On or about (b)(6) 2015, plaintiff underwent placement of an angiodynamics smartport ct device, catalog number h787ct80stpd0; lot number 4873601.The device was implanted by (b)(6), do at (b)(6) medical center in (b)(6), for the purpose of ongoing vein access for medication to treat plaintiff's loin pain-hematuria syndrome.The device was placed in plaintiff's right-sided internal jugular vein and right-side anterior chest wall.On or about (b)(6) 2015, the plaintiff was admitted to (b)(6) medical center in (b)(6), for evaluation and treatment of central venous catheter infection and bacteremia.Her line cultures and blood cultures were positive for both gram-negatives rods and cocci.Her treating physicians decided to remove the port-catheter device, which was performed on (b)(6) 2015 by (b)(6) do.Plaintiff remained inpatient until (b)(6) 2015, and she was treated with seven (7) days of intravenous antibiotics during her hospital stay.She was discharged with another seven (7) day course of oral antibiotics.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer's reported complaint description of patient infection cannot be confirmed given the patient centric nature of this serious adverse event (sae).No port device was returned for evaluation since there was no report of port device malfunction, damage or performance issue during in situ use/treatment.A device history record (dhr) review of the indicated packaging lot revealed no quality related issues or manufacturing deficiencies at the time of manufacture.Certificate of conformance for the sterilization process for the reported packaging lot was verified to be approved at time of distribution.The port was implanted into the patient 4 months prior to the patient's infection.There was no report of port device malfunction or performance issue during the implant procedure/use.If the manufacturing/packaging/sterilization of the port device in question was to contribute to an infection in the patient it would have likely occurred shortly after the implant date and not 4 months later.Device directions for use (dfu) cautions that each access of the port should be performed using aseptic technique, following the institutions universal precautions.There is no indication from the reported complaint that the manufacturing/packaging/sterilization of the port device could have contributed to the patient's infection.Infection is cautioned in the device dfu as a potential complication of port system use.Labeling review:
the directions for use (dfu) that is provided with the port device contains the following statements:
warnings:
· the device is to be implanted, used, maintained, and removed in strict accordance with institutional and or centers for disease control (cdc) guidelines or policies.· reuse of single-use devices creates a potential risk of patient or user infections.Contamination of the device may lead to injury, illness or death of the patient.· do not use syringes smaller than 10 ml syringe when accessing the port as system damage can occur.Flushing occluded catheters with small syringes can create excessive pressures within the port system.Precautions
· carefully read and follow all instructions prior to use.· strict aseptic technique is of paramount importance when implanting any device.· for peripheral placement, irritation to the vein, resulting in postoperative thrombophlebitis, has been associated with guidewire and introducer insertion.· when using percutaneous introducers: to avoid blood vessel damage, do not allow the percutaneous introducer sheath to remain indwelling in the blood vessel without the internal support of a catheter or dilator.· if more than one drug is to be administered, between the individual drug applications, flush the system with 5 to 10 ml normal saline for injection to prevent drug interactions.· after any infusion, injection or bolus application, the system should be flushed with normal saline for injection or locked with a heparin solution per institutional protocol to prevent thrombotic occlusion of the catheter.Potential complications
use of an angiodynamics port system involves potential risks normally associated with the insertion or use of any implanted device or indwelling catheter including but not limited to:
air embolism
catheter disconnection or migration
catheter embolization
catheter fragmentation
catheter pinch-off
clot formation
drug extravasation (leakage)
erosion of vessel and skin
implant rejection
infection
inflammation
thromboembolism
thrombophlebitis
thrombosis
necrosis of scarring of skin over implant area
vessel trauma
post-operative care
the incision site should be monitored for signs of infection, inflammation, hematoma, device rotation or erosion.Routine wound care should be given to these sites.The smart port ct implantable port may be used immediately after verification of catheter placement.Instruct patient to avoid any heavy exertion or strenuous activity during the first few days after surgery.General guidelines
· each access of an angiodynamics smart port ct implantable port should be performed using aseptic technique.· follow institutional universal precautions.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint continues to be monitored for trends.Reference (b)(4).
|
| |
|