Catalog Number 508-36-107
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Joint Laxity (4526)
Date of Event 07/09/2026
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery due to dislocation.
 
Additional Manufacturer Narrative
The agent reported, "dr (b)(6) had a patient he inherited from dr (b)(6) at (b)(6) hospital.This patient had existing lima humerus and djo glenoid.The patient dislocated.Dr (b)(6) lateralized his glenoid and lengthened the humerus by adding an extension and constrained poly.The patient left surgery very stable.Age 76 gender male." the previous surgery and the surgery detailed in this event occurred 1 month, 13 days apart.Item number: 508-36-107."item description: altivate reverse glenosphere 36mm/+4mm w/retaining screw." lot number: 4478a1238.Part revision: b.Product type: shoulder.Manufacture date: 24-feb-2026.Expiration date: 18-feb-2031.Note: lima-product reported by the customer will be investigated through lima.This evaluation is limited in scope because the reported device was not returned to enovis surgical for evaluation.The surgery was completed as intended and without reported complications.If additional information becomes available, this investigation may be reopened and reassessed.No patient activity, traumatic event, accident, or medical contraindication was reported that may have contributed to the event.There were no findings during this investigation indicating that the reported enovis device was defective.The revision surgery was performed to address the patient's instability/dislocation.Surgical technique / ifu review: the reported construct included implants from multiple manufacturers (lima humeral implant and djo/enovis glenoid component).Enovis does not recommend off-label use involving the combination of enovis implants with implants from other manufacturers unless specifically indicated in the applicable labeling and instructions for use.Based on the information provided, it could not be confirmed whether the mixed-manufacturer construct contributed to the reported instability.Compatibility assessment: the reported shoulder reconstruction consisted of a lima humeral component used in conjunction with a djo/enovis glenoid component.Because the construct involved implants from different manufacturers, compatibility between the components cannot be fully assessed through this complaint investigation.Root cause: the root cause of this complaint was a revision surgery due to dislocation of the shoulder arthroplasty requiring glenoid lateralization, humeral lengthening, and implantation of a constrained polyethylene liner.No information was submitted with the complaint regarding pre-existing patient conditions or patient activities that may have contributed to the event.Additionally, the reported construct consisted of a lima humeral implant used in combination with a djo/enovis glenoid implant.The potential impact of combining implants from different manufacturers cannot be ruled out and may have contributed to the reported issue.There are multiple factors that may contribute to an event that are outside the control of enovis surgical.Containment: inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.The revision surgery was completed successfully.Device history records review: a review of the device history record(s) shows that the reported component(s) used in the previous surgery met design and manufacturing requirements at the time of release for use.There were no nonconforming material reports associated with the product(s) that may have contributed to the reported event.The device(s) was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Complaint history: customer complaint history of the reported device(s) showed no present trends or ongoing issues that need review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DJO SURGICAL
Common Device Name
ALTIVATE REVERSE GLENOSPHERE 36mm/+4mm W/RETAINING SCREW
Manufacturer (Section D)
Encore Medical L.P
9800 metric blvd.
austin TX 78758
Manufacturer (Section G)
Encore Medical L.P
9800 metric blvd.
austin TX 78758
Manufacturer Contact
stacey bonnell
9800 metric blvd.
austin, TX 78758
MDR Report Key50064071
Report Number1644408-2026-01500
Device Sequence Number10121537
Product Code KWS
UDI-Device Identifier00190446759904
UDI-Public00190446759904
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K233481
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number508-36-107
Device Lot Number4478A1238
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/09/2026
Initial Report FDA Received Date07/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/24/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
Unknown Lot. 2227986.; Unknown Lot. 2541490.; Unknown Lot. 25AT70.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age76 YR
Patient SexMale
-
-