On (b)(6) 2026, a patient was scheduled to undergo a localization wire placement procedure using a ghiatas breast localization wire.Prior to the procedure, foreign material was observed on the needle.The procedure was subsequently completed using an alternative device.There was no patient contact.
|
|
H11: manufacturing review: the device history records have been reviewed, and this lot met all release criteria.Investigation summary: the physical device was not returned for evaluation.Two electronic photos were provided and reviewed.The first photo shows that a labeling information of the ghiatas which verifies and matches with tw details.The second photo shows that a wire clip and is expected to be positioned on the needle protector packaging.Based on the photo review the reported event cannot be confirmed.Therefore, the investigation is unconfirmed for the reported device contamination with chemical or other material, as no foreign material was observed in the provided photo.A definitive root cause for the reported device contamination with chemical or other materials could not be determined based upon the provided information
labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.D4 (unique identifier (udi) #), g3, h6 (method, result, conclusion)
section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|