| Model Number |
IDSS-SCC |
| Medical Device Problem Code |
Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/29/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The user facility reported that the monitors to their bdv idss custom room rack intermittently go black and reboot.This event did not occur during a patient procedure.No report of injury.
|
| |
|
Additional Manufacturer Narrative
|
|
A steris service technician arrived onsite following the reported event to inspect the unit and found that the power cable connection to the monitor was damaged causing intermittent power loss.To resolve the issue, the technician repaired the power cabling.The unit was tested, confirmed to be operating according to specification, and returned to service.No additional issues have been reported.
|
| |
|