The complaint source reported that a first-stage shoulder revision surgery was performed on (b)(6) 2026 due to a dislocation.During the procedure, the glenosphere, custom screw, liner, short reverse body, and associated screw were explanted.A new screw from a standard reverse body was inserted into the stem, which was left in situ.According to the surgeon, a possible contributing factor was an infection.The surgeon also noted that the reverse body screw was not fully engaged and that insufficient bone was removed from the area at reverse body / stem connection.A new custom screw has been requested to limacorporate for the planned second-stage revision surgery.Hereby the list of devices that were implanted in the previous surgery:
smr reverse humeral body short (product code: 1352.15.005, lot: 2602989 - ster: 260004) - sold in us
smr reverse liner standard (product code: 1360.50.010, lot: 24at2c8 - ster: 2400162) - not sold in us
smr eccent.Glenosphere ø 36mm (product code: 1376.15.031, lot: 2601926 - ster:2600023) - not sold in us
cortic.Bone screw d.5 l.32 mm (product code: 8432.15.032, lot: 2521662 - ster: 2500150) - sold in us
cortic.Bone screw d.5 l.34 mm (product code: 8432.15.034, lot: 2528462 - ster: 2500190) - sold in us
cortic.Bone screw d.5 l.36 mm (product code: 8432.15.036, lot: 2517337 - ster: 2500120) - sold in us
cortic.Bone screw d.5 l.48 mm (product code: 8432.15.048, lot: 2305906 - ster: 2300058) - sold in us
cmd* 25-1748 glenoid implant (product code: 9617.Pe0.A0, lot: 2610182 - ster: 2600074) - not sold in us
cmd* 25-1748 glenoid psi set 3 (product code: 9701.66.Azj, lot: 2610183 - ster: 2600060) - not sold in us
cmd* 25-1748 glenoid anatomy (product code: 9705.An.2ey, lot: 2610184 - ster: 2600060) - not sold in us
(*cmd: custom made device)
event occurre in united kingdom.
|