| Catalog Number |
68-360 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/02/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported "deflation".Healthcare professional later reported "any creases, any creases associated with hole and hole in radius (edge)".This record is for the left side.Device was explanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "deflation".Healthcare professional later reported "any creases, any creases associated with hole and hole in radius (edge)".This record is for the left side.Device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
This report was impacted by emdr scheduled maintenance occurring july 22-27, 2026.A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected: d9, h3 and h6.Continued e1: alternate email address: (b)(6).Device evaluation:
per the investigation procedure, the device is analyzed through visual inspection microscopic inspection if openings are observed and a weight verification.Per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required:
¿ deflation: observed two openings through microscopic inspection 1 opening assessed as surgical damage and 1 opening assessed as fold flaw opening.No further actions are required as it is not related to the manufacturing process.¿ crease / folding of implant: observed creases on device through visual inspection.No further actions are required as it is not related to the manufacturing process.None of the other observations performed during the device analysis wear abrasion are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
|
| |
|