It was reported pericardial effusion and cardiac tamponade occurred.A left atrial appendage closure (laac) procedure was being performed.A watchman fxd double curve access system (was) was selected for use.The initial transeptal puncture (tsp) was performed using a non-boston scientific (bsc) needle and intracardiac echo (ice) imaging.Tsp was difficult to visualize tenting and positioning of the needle on the septum.Following tsp, the physician advanced the was sheath.Several minutes later, the patient experienced a pericardial effusion followed by cardiac tamponade.A pericardiocentesis was performed to treat the effusion, draining approximately 2200 ml of blood, which was transfused back to the patient.The physician was able to successfully clip the appendage, and the laac implant procedure was cancelled due to emergent need to stabilize the patient.The patient remained in the hospital for several days while they recover.
|