| Model Number |
72404452 |
| Medical Device Problem Codes |
Collapse (1099); Fluid/Blood Leak (1250); Inflation Problem (1310); Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/01/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient with an inflatable penile prosthesis (ipp) was unable to inflate the device.The pump remained compressed and flat due to fluid loss caused by a break in the pump connecting tubing.A surgical procedure was performed, during which the ipp was removed and replaced with a tactra malleable device at the patient's request due to manual dexterity issues.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Model number/catalog number: (b)(4)
serial number: n/a
batch/lot number: (b)(4)
model/catalog description: (b)(4)
unique identifier (udi) #: (b)(6).
|
| |
|