| Model Number |
9-ACP2-007-022 |
| Medical Device Problem Code |
Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Codes |
Angina (1710); Cardiac Arrest (1762); Perforation (2001); Cardiac Tamponade (2226); Pericardial Effusion (3271)
|
| Date of Event |
07/01/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported on 29 june 2026 a 22 mm amplatzer amulet left atrial appendage occluder was selected for implant using a 12f amplatzer amulet delivery sheath for a left atrial appendage (laa) closure.During the procedure, the left atrial pressure was above 12 mmhg.The activating clotting factor (act) was 339 seconds and heparin was administered.The heart rhythm was atrial fibrillation.Transseptal puncture was made at the inferior and mid location.Protamine 30 cc was administered.The device was implanted.On (b)(6) 2026, the patient experienced chest pain and went into cardiac arrest.Cardiac tamponade was determined to be present.Injury to the pulmonary artery was noted from the stabilizing wires of the device.Surgical intervention was performed using a surgical patch on the pulmonary artery to resolve the issue.The patient status was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
The device remains implanted in patient.The device will not be returned for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|