DePuy Ireland - 9616671 VELYS; Orthopedic Manual Surgical Instrument Back to Search Results
Catalog Number 451570005
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Limb Fracture (4518)
Date of Event 01/01/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the surgeon believes fracture of a distal femur may have been caused by bicortical velys array pins causing stress riser.He mentioned that he would plan to place pins more distal and unicortical as per velys ras standard pin placement.One patient treated non operatively, one patient treated with retrograde nail.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11.Additional manufacturer narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VELYS
Common Device Name
Orthopedic Manual Surgical Instrument
Manufacturer (Section D)
DePuy Ireland - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
+1(908)386-3534
MDR Report Key50071492
Report Number1818910-2026-12912
Device Sequence Number18672375
Product Code LXH
UDI-Device Identifier10603295519096
UDI-Public(01)10603295519096
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
Exempt
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number451570005
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 07/05/2026
Supplement Date Received by Manufacturer08/09/2026
Initial Report FDA Received Date07/29/2026
Supplement Report FDA Received Date08/10/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-