| Model Number |
3500-536U3 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
Joint Dislocation (2374)
|
| Date of Event |
07/02/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A patient who had received a custom proximal humerus reconstruction reported experiencing a dislocation during physical therapy.The physician was not able to perform a closed reduction of the shoulder, so performed a surgery to correct it.It was discovered intraoperatively that the locking mechanism of the polymer liner had been damaged and had disassociated from the humeral implant.A new liner was implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Review of production records associated with the manufacture, packaging, and sterilization of the devices in question reveals no abnormalities.Humeral component dissociations are a common issue in reverse shoulder arthroplasty.
|
| |
|