The product investigation was completed.Device evaluation details:
the device was returned to johnson & johnson medtech (j&j medtech) for evaluation.Following j&j medtech procedures, a visual inspection and revision of all device features were performed.Visual inspection revealed no damage or anomalies on the device.The device was connected to carto 3 system and an ablation cycle was performed, no force, magnetic, noise, temperature or impedance issues were found.Additionally, device was deflected and irrigated during the test and no issues were observed.A manufacturing record evaluation was performed for the finished device lot number: 31535591l and no internal action was found during the review.No malfunction was observed during the product analysis.The root cause of the adverse event remains unknown.The instructions for use (ifu) states the following recommendations: when using the catheter with conventional systems (such as fluoroscopy or intracardiac echocardiography), with the carto¿ 3 system, careful catheter manipulation must be performed in order to avoid cardiac damage, perforation, or tamponade.Catheter advancement should be done under direct imaging guidance.Do not use excessive force to advance or withdraw the catheter when resistance is encountered.The firmness of the braided tip dictates that care must be taken to prevent perforation of the heart.As part of johnson & johnson medtech's quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|