• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DePuy Ireland - 3015516266 SMARTSET; Bone Cement : Bone Cement

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DePuy Ireland - 3015516266 SMARTSET; Bone Cement : Bone Cement Back to Search Results
Catalog Number 3102080
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 07/08/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient experienced aseptic tibial loosening.Tibia loosened at the cement/implant interface.All components were revised.Femur was not loose.No patient injury or delay.Doi: unknown.Dor: on (b)(6) 2026.Affected side: right knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional narrative: h3, h6: investigation summary: no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not performed.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: a previous device history record (dhr) review ((b)(4)) was performed.The dhr review revealed of the (b)(4) unit batch, four unrelated non-conformances of this lot were identified.Final micro and sterility tests passed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMARTSET
Common Device Name
Bone Cement : Bone Cement
Manufacturer (Section D)
DePuy Ireland - 3015516266
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
CMW Mfg & Material Warehouse
UK  
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
+1(908)386-3534
MDR Report Key50073333
Report Number1818910-2026-12936
Device Sequence Number10480455
Product Code LOD
UDI-Device Identifier10603295168355
UDI-Public(01)10603295168355(17)161129(10)7786781
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K042591
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/29/2016
Device Catalogue Number3102080
Device Lot Number7786781
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 07/08/2026
Initial Report FDA Received Date07/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/04/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE PS FB INSRT SZ 4 6MM; ATTUNE PS FEM RT SZ 4 CEM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age67 YR
Patient SexFemale
-
-