The device was not returned for analysis.A cine review was performed on the imaging that confirmed what appears to be a broken wire segment in the artery.Production record and corrective and preventive actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a query of the complaint handling database for the reported lot revealed there is no indication of a lot specific issue.Based on the information provided, a definitive cause for the reported issue could not be determined.Factors that may contribute to material separation include, but are not limited to, tensile strength, corrosion, force applied to device, interaction with accessory devices, and/or challenging anatomy.Based on the results of the complaint investigation, there is no indication of a product quality issue with respect to manufacture, design or labeling of the device.
|