• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Centinel Spine, LLC PRODISC C VIVO US IMPLANT LARGE 5mm; PROSTHESIS, INTERVERTEBRAL DISC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Centinel Spine, LLC PRODISC C VIVO US IMPLANT LARGE 5mm; PROSTHESIS, INTERVERTEBRAL DISC Back to Search Results
Model Number PDVL5
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Neck Pain (2433)
Date of Event 07/09/2026
Type of Reportable Event Serious Injury
Event or Problem Description
A patient was implanted with a prodisc c vivo device at c5-6 on (b)(6) 2025.The patient was experiencing neck pain.It was found that the superior endplate had migrated anteriorly.The implant was removed on (b)(6) 2026 and the patient was fused.
 
Additional Manufacturer Narrative
An mdr is indicated for this complaint.A patient was implanted with a prodisc c vivo device at c5-6 on (b)(6) 2025.The patient was experiencing neck pain.It was found that the superior endplate had migrated anteriorly.The implant was removed on (b)(6) 2026 and the patient was fused.Per the distributor, the surgeon determined that the posterior disc space was not sufficiently opened causing a slope and contributing to the migration of the device.A dhr review found no anomalies associated with the complaint.Complaint trending found that the rate of complaints is within the level outlined in the risk documentation.A review of the risk documentation found that the hazards associated with the complaint are identified and mitigated to a level where the clinical benefits outweigh the harms.The implant was returned following the removal surgery and will be evaluated using an approved cervical implant evaluation protocol.Imaging was provided that showed the device had migrated.There were no anomalies identified during the complaint investigation.The removal was completed due to implant migration.The submission is 1 of 1 devices involved in this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRODISC C VIVO US IMPLANT LARGE 5mm
Common Device Name
PROSTHESIS, INTERVERTEBRAL DISC
Manufacturer (Section D)
Centinel Spine, LLC
900 airport road
suite 3b
west chester PA 19380
Manufacturer (Section G)
Fruh Exton
240 sierra drive
suite 100
exton PA 19341
Manufacturer Contact
jasmine carrillo
900 airport road
suite 3b
west chester, PA 19380
+1(215)850-0881
MDR Report Key50073418
Report Number3007494564-2026-00068
Device Sequence Number10206776
Product Code MJO
UDI-Device Identifier00843193113733
UDI-Public(01)00843193113733(17)291130(10)2024-1225
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P070001
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberPDVL5
Device Catalogue NumberN/A
Device Lot Number2024-1225
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/28/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 07/02/2026
Initial Report FDA Received Date07/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age44 YR
Patient SexMale
-
-