| Model Number |
M635WU50270 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/27/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported device migration occurred.A left atrial appendage (laa) closure procedure was performed, and a 27mm watchman flx closure device was implanted on (b)(6)-2023.The patient was discharged.At a routine follow up, it was found that the closure device had shifted very proximal in the appendage and was only attached by a few of the bottom row anchors.The physician removed the closure device using a percutaneous snare and implanted a new 27mm watchman flx pro closure device without complications.
|
| |
|
Event or Problem Description
|
|
It was reported device migration occurred.A left atrial appendage (laa) closure procedure was performed, and a 27mm watchman flx closure device was implanted on (b)(6)2023.The patient was discharged.At a routine follow up, it was found that the closure device had shifted very proximal in the appendage and was only attached by a few of the bottom row anchors.The physician removed the closure device using a percutaneous snare and implanted a new 27mm watchman flx pro closure device without complications.
|
| |
|
Additional Manufacturer Narrative
|
|
With all the available information, boston scientific (bsc) concludes:
device technical analysis
the watchman flx? was not returned; therefore, device analysis could not be completed.Media review
the provided medical media was captured during a follow up appointment and does not require further review by either bsc medical safety or watchman medical media subject matter experts.Device history record review
a review was completed of the device history records (dhr) for the watchman flx?, part # m635wu50270 batch # 0030624765.The dhr review confirmed that the device met all material, assembly, and performance specifications prior to distribution.Labeling review
there was no evidence to suggest that the device was used in a manner inconsistent with the labeling.Watchman flx? instructions for use (ifu) lists potential complications, risks and adverse events that may be associated with the use of this device which include implant movement/shift.Risk review
a risk review was completed and confirmed that the event of implant movement/shift was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Investigation conclusion
based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.
|
| |
|