MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. Visia AF MRI XT VR SureScan; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. Visia AF MRI XT VR SureScan; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) Back to Search Results
Model Number DVFB1D4
Medical Device Problem Codes Signal Artifact/Noise (1036); Electromagnetic Interference (1194); Inappropriate/Inadequate Shock/Stimulation (1574); Device Sensing Problem (2917)
Health Effect - Clinical Codes Non specific EKG/ECG Changes (1817); Shock from Patient Lead(s) (3162)
Date of Event 07/07/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that the patient had a procedure using cautery.Prior to the procedure, there were no issues with implantable cardioverter defibrillator (icd) pacing or the patient's heart rate.However, during the procedure the patient's heart rate was interrupted.A magnet was not placed on the patient's icd at the beginning of the procedure.The magnet was placed over an hour into the procedure.At the time the magnet was placed, the patient's heart rate went to forty beats per minute and the device's lower rate was set to that.During the procedure, prior to the magnet being placed, a lead integrity alert (lia) triggered for high rate non-sustained episodes and short ventricular intervals.An abrupt increase in right ventricular (rv) lead impedance was noted along with oversensing noise on the triggering misclassification of events and inappropriate high-voltage therapy.The icd remains in use.No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Visia AF MRI XT VR SureScan
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112--439
+1(763)526-9277
MDR Report Key50074826
Report Number3004209178-2026-12727
Device Sequence Number10744043
Product Code LWS
UDI-Device Identifier00643169720794
UDI-Public(01)00643169720794(17)210514
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P980016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/31/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date05/14/2021
Device Model NumberDVFB1D4
Device Catalogue NumberDVFB1D4
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/07/2026
Supplement Date Received by Manufacturer08/27/2026
Initial Report FDA Received Date07/29/2026
Supplement Report FDA Received Date08/31/2026
Date Device Manufactured11/21/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
6935M55 LEAD
Outcome Attributed to Adverse Event Life Threatening;
Patient Age62 YR
Patient SexFemale
Patient Weight115 LB
-
-