It was reported that a patient experienced a return of pain in the setting of high impedance.The high impedance was not observed on the day of surgery, and impedance values were referenced as greater than 40 ,000.Programming was changed to working electrodes.Provider will order x-ray to verify lead placement. the issue was reported as ongoing.The manufacturer representative (rep) indicated they would send any further information as they become aware.2026-jul-07 mpxr 1448886.1 (rep): additional information was received from the manufacturer representative (rep).It was reported that the rep noted the top of the right lead has migrated down to t9.2026-jul-13 mpxr 1448886.2 (rep): additional information was received from the manufacturer representative (rep).The same rep later noted conflicting information that they met with patient today with return of pain.X-rays were taken and leads are in the same position as implant day.High impedances are noted on electrodes 1,2,3,4,6,12,13,14.Rep programmed around these electrodes and the patient will updated me at a later date with pain results.
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
|