| Model Number |
352506070E |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
06/28/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Customer reported during ivc filter placement, the physician encountered difficulty removing the dilator from the sheath.When he pulled again on the dilator, it came out but the radiopaque markers were not on the dilator.Fluoroscopy showed one marker left behind in the patient.The physician was able to remove the marker from the patient.They opened a different kit and completed the procedure successfully.Only the dilator is available for return.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
The complaint sample has been returned for investigation and is currently in process.A follow up report will be submitted upon the closure of the investigation.
|
| |
|
Search Alerts/Recalls
|