• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Curonix LLC Freedom Peripheral Nerve Stimulator

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Curonix LLC Freedom Peripheral Nerve Stimulator Back to Search Results
Model Number STQ4-RCV-A0
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Impaired Healing (2378)
Date of Event 07/07/2026
Type of Reportable Event Serious Injury
Event or Problem Description
The patient reported their receiver site had not healed.Cultures were obtained.However, an infection was not confirmed.The patient has been going to wound care for three months where they have been monitoring, debriding, and dressing the wound.Additionally, the patient completed a short term steroid to improve healing.As of (b)(6) 2026, the incision has improved significantly since their previous visit and the wound has reduced in size greater than 50%.The patient will follow up with wound clinic, they are using their device and getting great relief.No further issues have been reported.
 
Additional Manufacturer Narrative
The surgical issue questionnaire was reviewed for potential causes of the reported issue.Based on this review, implanting the device after the expiration date, not prepping the surface of the skin with an antiseptic solution, not irrigating the incision site with an antibiotic solution before closure, multiple tunneling attempts, and using incorrect tools have been ruled out.Additionally, improperly closing the surgical site, the patient having contraindicating conditions, and the patient picking or touching the wound have been ruled out.A curonix representative conducted a review of sterilization and packaging records for the respective product lot; curonix has confirmed that the product was delivered sterile, validated sterilization parameters were used, and sterile barriers were verified to be intact following packaging.The stimulator is used to treat pain.The cause of the impaired healing is unknown.The investigation findings do not lead to a clear conclusion about the cause of the reported issue.Therefore, the conclusion has been selected as unable to determine the root cause.Rates reviewed at the most recent complaint trending meeting do not indicate a significant increase in surgical issues.Surgical issue rates remain acceptably low; thus, a capa is not required at this time.Surgical issue rates will continue to be tracked and trended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Freedom Peripheral Nerve Stimulator
Common Device Name
Peripheral Nerve Stimulator
Manufacturer (Section D)
Curonix LLC
1310 park central boulevard s.
pompano beach FL 33064
Manufacturer (Section G)
Curonix LLC
1310 park central boulevard s.
pompano beach FL 33064
Manufacturer Contact
alicia pagliaro
1310 park central boulevard s.
pompano beach, FL 33064
+1(800)965-5134
MDR Report Key50076670
Report Number3010676138-2026-00146
Device Sequence Number10744657
Product Code GZF
UDI-Device Identifier00850051034406
UDI-Public(01)00850051034406(17)280201
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K233162
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberSTQ4-RCV-A0
Device Lot NumberSWO260203
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/07/2026
Initial Report FDA Received Date07/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/27/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age59 YR
Patient SexFemale
-
-