• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. Azure XT SR MRI SureScan; PULSE GENERATOR, PERMANENT, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. Azure XT SR MRI SureScan; PULSE GENERATOR, PERMANENT, IMPLANTABLE Back to Search Results
Model Number W1SR01
Medical Device Problem Code Pacing Problem (1439)
Health Effect - Clinical Codes Bradycardia (1751); Non specific EKG/ECG Changes (1817)
Date of Event 03/24/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the right atrial (ra) lead, which is place in the right ventricle port, exhibited a possible sensing issue on presenting electrograms (egm) and short v-v intervals.The ra lead remains un use.It was further reported that the lead exhibited intermittent undersensing on current electrograms (egm) with false classification/te rmination of ongoing arrhythmias.It was further reported that the patient experienced sinus pause for two minutes with a heart rate in the fifties.Gaps were also observed between the qrs on external cardiac monitor.The ra lead exhibited undersensing of atrial fibrillation (af) waves noted resulting in unnecessary pacing at times.The implantable pulse generator (ipg) remains in use.No further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Azure XT SR MRI SureScan
Common Device Name
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112--439
+1(763)526-9277
MDR Report Key50077333
Report Number3004209178-2026-12357
Device Sequence Number18722128
Product Code NVZ
UDI-Device Identifier00643169634626
UDI-Public(01)00643169634626(17)191028
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P980035
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/28/2019
Device Model NumberW1SR01
Device Catalogue NumberW1SR01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/06/2026
Supplement Date Received by Manufacturer09/26/2026
Initial Report FDA Received Date07/29/2026
Supplement Report FDA Received Date09/30/2026
Date Device Manufactured05/07/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
5076-45 LEAD
Outcome Attributed to Adverse Event Life Threatening;
Patient Age83 YR
Patient SexMale
Patient Weight123 LB
-
-