• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medtronic Puerto Rico Villalba Sprint Quattro Secure S MRI SureScan; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medtronic Puerto Rico Villalba Sprint Quattro Secure S MRI SureScan; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) Back to Search Results
Model Number 6935M55
Medical Device Problem Codes Signal Artifact/Noise (1036); Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574)
Health Effect - Clinical Code Shock from Patient Lead(s) (3162)
Date of Event 07/11/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the right ventricular (rv) lead triggered a lead integrity alert (lia) due to sensing integrity counter (sic) and high rate non-sustained episodes.In addition both the rv and right atrial (ra) lead exhibited noise and oversensing.The patient received an inappropriate shock due to the noise and oversensing.It was also noted that the patient was in a swimming pool at the time of the lia.The rv and ra leads remain in use.No further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Continuation of d10: dtpa2qq crt-d, implanted: (b)(6) 2023.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Sprint Quattro Secure S MRI SureScan
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
Medtronic Puerto Rico Villalba
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
Medtronic Puerto Rico Villalba
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112--439
+1(763)526-9277
MDR Report Key50077685
Report Number2649622-2026-22262
Device Sequence Number10760695
Product Code LWS
UDI-Device Identifier00643169309760
UDI-Public(01)00643169309760(17)170220
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
P920015
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date02/20/2017
Device Model Number6935M55
Device Catalogue Number6935M55
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/12/2026
Initial Report FDA Received Date07/29/2026
Date Device Manufactured02/25/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
459878 LEAD, 5076-45 LEAD
Outcome Attributed to Adverse Event Life Threatening;
Patient Age68 YR
Patient SexFemale
Patient Weight133 LB
-
-