| Model Number |
D-EVOLUTFX-2329 |
| Medical Device Problem Codes |
Failure to Advance (2524); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
Vascular Dissection (3160)
|
| Date of Event |
07/28/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Additional information was received indicating that the injury occurred during insertion and advancement.It was reported that the non-medtronic sheath was passed through the stenosed/dissected area and that the delivery catheter system (dcs) never passed though it.Dissection was not confirmed.Per the physician, the cause of the injury was unknown.It was noted that the minimum diameter of the access vessel was 5 millimeter (mm) x 6mm.
|
| |
|
Event or Problem Description
|
|
It was reported that during a transcatheter aortic valve implant procedure, the delivery catheter system (dcs) was inserted into the patient's right femoral artery.There was then difficulty in advancing the catheter within the patient's right femoral artery.Subsequently, a balloon angioplasty was completed of the right femoral artery and external iliac artery.A possible dissection was then noted.Per the physician, the patient's calcified anatomy contributed to the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated: b5 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Continuation of d10: product id evfxplus-26 (serial: (b)(6)); product type: 0195-heart valves; implant date: n/a ; explant date: n/a medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|