| Catalog Number |
0112980 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Codes |
Adhesion(s) (1695); Hernia (2240)
|
| Date of Event |
03/17/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Per information provided:
on (b)(6) 2013 - patient was diagnosed with recurrent ventral incisional hernia and left inguinal hernia thereby underwent open recurrent ventral incisional hernia repair with the implant of ventrio mesh (device 1) and left inguinal hernia repair with the implant of perfix plug (device 2).Per op notes, "a midline incision was made, ventrio mesh (device 1) was placed towards the abdominal cavity using prolene sutures.Attention to the left inguinal hernia, where a left incision was performed.A perfix plug (device 2) was placed to the abdomen using sutures."
on (b)(6) 2016 - patient was diagnosed with recurrent ventral incisional hernia hereby underwent laparoscopic repair with the removal of both meshes (device 1 and 2) and implant of ventrio st mesh (device 3).Per op notes, "patient had recurrent ventral incisional hernia which had been repaired previously with two meshes.Extensive intra-abdominal adhesions were taken down.The more recent mesh (device 1 and 2) which had been placed appeared to have folded over on itself.Both meshes (device 1 and 2) were removed from the abdominal wall.A ventrio st mesh (device 3) was secured to the old mesh inferiorly using sutures.".
|
| |
|
Additional Manufacturer Narrative
|
|
Root cause is inconclusive; no conclusion can be made as to the extent to which the implant may have caused or contributed to the patient's postoperative course.Note, the manufacturing date (15-mar-2013) is considered to be a best estimate.This emdr represents the bard perfix plug (device 2).An additional supplemental emdr was submitted to represent the ventrio mesh (device 1).Note: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Search Alerts/Recalls
|