W. L. Gore & Associates, Inc. GORE® DrySeal Flex Introducer Sheath; Introducer, Catheter
|
Back to Search Results |
|
| Catalog Number |
DSF2233 |
| Medical Device Problem Code |
Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
Vascular Dissection (3160)
|
| Date of Event |
07/13/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The following was reported to gore:
on (b)(6), 2026, the patient underwent endovascular treatment of a residual dissection in the thoracic aorta following total arch replacement and open stent grafting.Two gore® tag® conformable thoracic stent graft with active control system devices were implanted using a 22fr gore® dryseal flex introducer sheath.After removal of the 22fr sheath, the left peripheral pulse was not palpable.Angiography, ultrasound, and ivus revealed a dissection extending from the left external iliac artery to the left common femoral artery (puncture site).Reportedly, there was no significant difference in blood pressure between the right and left sides.Therefore, the physician decided to monitor the patient.It was reported that the vessel diameter from the left common femoral artery to the left external iliac artery ranged from approximately 6.2 to 9.5 mm.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: a review of the manufacturing records for the device verified that the lot met all pre-release specifications.This complaint includes information indicating the reported complications relate to use of a sheath size that was too large for the patient access vessel anatomy per the device instructions for use.The device instructions for use provide warning and instruction to ensure the vessel diameter is adequate to accommodate the device.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Search Alerts/Recalls
|
|
|