Model Number 0684-00-0576-01U
Medical Device Problem Codes Deflation Problem (1149); Difficult to Insert (1316); Suction Problem (2170)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/09/2026
Type of Reportable Event Malfunction
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
Updated fields: b4, b6, b7, d10, d9, g3, g6, g7, h2, h3, h4, h6 (medical device problem code, component codes, type of investigation, investigation findings and investigation conclusions), h11.Corrected fields: d4 serial (revert this field to blank, as it was inadvertently submitted initially) and d8 (revert this field to blank, as it was inadvertently submitted initially).The iab was returned with the membrane completely unfolded and blood on the exterior of the catheter.The one-way valve was also returned.The sheath was not returned for evaluation.The one way valve was vacuum tested and it held vacuum.A laboratory insertion test was unable to be performed due to the membrane being unfurled.An underwater leak test of the balloon, catheter, y fitting and extracorporeal tubing was performed and no leaks were detected.A leak may impact the ability to maintain vacuum.Difficult or unable to draw vacuum on iab was unable to be confirmed.Additionally we are unable to confirm the reported difficulty during insertion because of the returned condition of the catheter and we are unable to mimic the clinical setting.Per ifu the one way valve must be aspirated to 30cc while leaving the one way valve in place attached to the extracorporeal tubing leur to ensure vacuum is maintained throughout insertion.The failure mode is addressed in the risk file and is operating within its risk profile.The ifu addresses the reported failure.The investigation does not indicate that the device was inadvertently released as non conforming or an adulterated product or was a counterfeit.The complaint history review did not identify an adverse trend.Based on the rational provided above, no escalation to the capa process is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Sensation Plus 8Fr. 50cc IAB with Accessories, US Only
Common Device Name
System, Balloon, Intra-Aortic And Control
Manufacturer (Section D)
Datascope Corp.
15 law drive
fairfield NJ 07004
Manufacturer (Section G)
Datascope Corp.
15 law drive
fairfield NJ 07004
Manufacturer Contact
ewelina chorlowska
15 law drive
fairfield, NJ 07004
MDR Report Key50079529
Report Number2248146-2026-0001344
Device Sequence Number13316735
Product Code DSP
UDI-Device Identifier10607567109633
UDI-Public(01)10607567109633(10)3000511517(17)281105(11)251105
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K133074
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0684-00-0576-01U
Device Catalogue Number0684-00-0576-01U
Device Lot Number3000511517
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/30/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/10/2026
Supplement Date Received by Manufacturer09/22/2026
Initial Report FDA Received Date07/30/2026
Supplement Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/05/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
Unknown
Patient Age74 YR
Patient SexMale
-
-