| Model Number |
7053HP |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
06/29/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The user facility reported while an employee was cleaning their amsco 7053hp washer/disinfector the panel that cover's the facility's plumbing/water inlet adjacent to the unit detached and "struck" the employee on the arm.It is unknown if medical treatment was sought or administered.
|
| |
|
Additional Manufacturer Narrative
|
|
A steris service technician arrived onsite to inspect the amsco 7053hp washer/disinfector and confirmed the reported event.The technician was unable to determine a cause of the reported event.To resolve the issue the technician secured the panel and confirmed the positioning of the panel's latches.The technician confirmed the amsco 7053hp washer/disinfector to be operating according to specification and returned the unit to service.A 3-year complaint review indicates this to be an isolated event.No additional issues have been reported.
|
| |
|