• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DePuy Orthopaedics Inc US PINNACLE GRIPTION; Pinnacle Hip System : Hip Metal Acetabular Cups

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DePuy Orthopaedics Inc US PINNACLE GRIPTION; Pinnacle Hip System : Hip Metal Acetabular Cups Back to Search Results
Catalog Number 121732052
Medical Device Problem Codes Device Dislodged or Dislocated (2923); Noise, Audible (3273)
Health Effect - Clinical Code Pain (1994)
Date of Event 07/03/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that patient had a revision hip procedure through a direct lateral approach.The original hip replacement procedure with pinnacle trilock implants was completed 7 years previously.The patient had presented in ae with a two day history of pain, and reported hearing a ¿clunk¿ sound.The revision procedure involved the extraction of a pinnacle sector cup 52mm, a poly 52/32mm, and a 32mm ceramic head.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINNACLE GRIPTION
Common Device Name
Pinnacle Hip System : Hip Metal Acetabular Cups
Manufacturer (Section D)
DePuy Orthopaedics Inc US
700 orthopaedic drive
warsaw IN 46581 -098
Manufacturer (Section G)
JTE Warsaw Mfg Site
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
+1(908)386-3534
MDR Report Key50083533
Report Number1818910-2026-12982
Device Sequence Number13316776
Product Code KWA
UDI-Device Identifier10603295010302
UDI-Public(01)10603295010302(17)290430(10)9169682
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K231873
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number121732052
Device Lot Number9169682
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/30/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 07/10/2026
Supplement Date Received by Manufacturer07/30/2026
09/22/2026
Initial Report FDA Received Date07/30/2026
Supplement Report FDA Received Date07/31/2026
09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/22/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-