| Model Number |
SONATA |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Codes |
No Clinical Signs, Symptoms or Conditions (4582); Infectious Otitis Media (4815)
|
| Date of Event |
04/13/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The user went to an ent clinic for ear canal cleaning.The doctor by mistake sucked the electrode out of the cochlea.The active electrode now is in the eac.The user has been explanted.
|
| |
|
Event or Problem Description
|
|
The user went to an ent clinic for ear canal cleaning.The doctor by mistake sucked the electrode out of the cochlea.The active electrode now is in the eac.The user has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: according to the information received from the field the active electrode was accidentally pulled out of cochlea during an ear cleaning procedure.Reportedly during explantation surgery an infection in the middle ear was found, which was likely the cause for the ear cleaning.Review of the device_s sterilization records show that the device has been subject to a valid sterilization process, thus no available information points to the implant being the source of the reported issue.The concerned device was explanted but has not been received for investigation yet.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has been explanted and should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|