| Model Number |
104523 |
| Medical Device Problem Codes |
Failure to Charge (1085); Migration (4003)
|
| Health Effect - Clinical Code |
Inadequate Pain Relief (2388)
|
| Date of Event |
07/07/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A patient implanted with an evoke spinal cord stimulation (scs) system reported difficulty charging their evoke closed loop stimulator (cls).The patient reported a previous fall; the evoke scs system didn¿t cause or contribute to the patient¿s fall.X-ray imaging identified migration of two (2) leads and the cls was in an angled position.The physician elected to perform a revision procedure to replace the migrated leads and reposition the cls.During the revision procedure, the physician encountered scar tissue and could not regain posterior access to place the new leads.As a result, the revision procedure was aborted, the patient subsequently underwent explantation of the evoke scs system.
|
| |
|
Additional Manufacturer Narrative
|
|
Cls information:
brand name: evoke closed loop stimulator (cls) (us)
model: 102901
catalog: 3042
serial number: (b)(6)
udi: 09352307001572
expiration date: 21september2026.
|
| |
|