Saluda Medical Pty Ltd Direct Clamp Active Anchor Kit; SCS Anchor Back to Search Results
Model Number 104523
Medical Device Problem Codes Failure to Charge (1085); Migration (4003)
Health Effect - Clinical Code Inadequate Pain Relief (2388)
Date of Event 07/07/2026
Type of Reportable Event Serious Injury
Event or Problem Description
A patient implanted with an evoke spinal cord stimulation (scs) system reported difficulty charging their evoke closed loop stimulator (cls).The patient reported a previous fall; the evoke scs system didn¿t cause or contribute to the patient¿s fall.X-ray imaging identified migration of two (2) leads and the cls was in an angled position.The physician elected to perform a revision procedure to replace the migrated leads and reposition the cls.During the revision procedure, the physician encountered scar tissue and could not regain posterior access to place the new leads.As a result, the revision procedure was aborted, the patient subsequently underwent explantation of the evoke scs system.
 
Additional Manufacturer Narrative
Cls information: brand name: evoke closed loop stimulator (cls) (us) model: 102901 catalog: 3042 serial number: (b)(6) udi: 09352307001572 expiration date: 21september2026.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Direct Clamp Active Anchor Kit
Common Device Name
SCS Anchor
Manufacturer (Section D)
Saluda Medical Pty Ltd
5 eden park drive
macquarie park, nsw -
AU  -
Manufacturer (Section G)
Saluda Medical Pty Ltd
5 eden park drive
macquarie park, nsw -
AU   -
Manufacturer Contact
sadie mauser
5 eden park drive
macquarie park, nsw -
AU   -
MDR Report Key50084974
Report Number3016111136-2026-00154
Device Sequence Number10479751
Product Code LGW
UDI-Device Identifier09352307001817
UDI-Public09352307001817
Combination Product (Y/N)N
Initial Reporter StateWV
Initial Reporter CountryUS
PMA/510(K) Number
P190002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Importer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/30/2026
Report Date (Section F) 07/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/12/2026
Device Model Number104523
Device Catalogue Number3043
Device Lot Number9017991808
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date07/07/2026
Event Location Hospital
Date Report to Manufacturer07/07/2026
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date07/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexMale
-
-