Paragon 28 Gorilla Plating System MTP, 0 degree, Medium Plate, Left; Plate, Fixation, Bone
|
Back to Search Results |
|
| Model Number |
N/A |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
Pain (1994); Nonunion/Delayed-union Bone Fracture (2369)
|
| Date of Event |
06/24/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that an mtp plate broke in half, revision surgery took place approximately 15 months post implantation due to patient experiencing pain ¿ existing hardware removed.Used p28 hardware used to complete revision.Screws are fine.Patient was not involved in any car accident, slip trip or fall but patient did have a non-union.The plate broke at the joint.Attempts have been made and not available information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4)
complaint was confirmed.Visual evaluation was completed and found that the device was fractured.The device was submitted for further analysis.Fracture analysis has been previously analyzed on a zrm document where fast-brittle type tensile/bending overload was identified as the mode of failure.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.Review of complaint history found no additional related issues for this item and the reported part and lot combination.Medical records were not provided.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Search Alerts/Recalls
|
|
|