It was reported that this inflatable penile prosthesis (ipp) was displaced.The patient complained of a significant discrepancy between the right and left corporal cylinders, as well as difficulty operating the device.Magnetic resonance imaging (mri) was performed and demonstrated proximal erosion of the right cylinder extending posteriorly within the perineum beyond the crus of the corpora.The distal aspect of the left cylinder terminated approximately 2.0 cm proximal to the right cylinder and did not reach the glans.The patient experienced floppy glans.The mri also suggested possible buckling of both cylinders within the penile shaft.A surgical procedure was performed, during which the ipp was removed.No further patient complications were reported.
|
|
The device is not available for analysis; therefore, no physical analysis of the product could be performed.The reported device performance allegation cannot be confirmed.The device history record (dhr) review confirmed that the device met all material, assembly and performance specifications.A review of the device instructions for use (ifu) was completed and did not reveal any evidence of device misuse, off label use, or failure to follow instructions.A risk review was completed and confirmed that the events of disfigurement, erosion, incorrect cylinder size, device migration (pump/cylinder) and mechanical malfunction/revision due to mechanical malfunction were defined in the product risk documentation.This event type has been accounted for during analysis to support acceptable risk benefit for the product.Based on the information available the cause that contributed to the reported event cannot be established.Therefore, a conclusion code of cause not established was assigned to this investigation.
|