| Model Number |
30-103 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/30/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A patient underwent a thrombectomy with the artix system of devices on (b)(6) 2026.Mid procedure, the artix sheath separated at the junction of the hemostasis valve and became stuck in the body.The funnel catheter remained inside the sheath.This was noticed under fluoro imaging and confirmed by seeing the funnel catheter outside of body.Removing the device required bilateral cutdowns.The patient was then treated with a different intervention.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer reference number: (b)(4).The suspect device was not returned to the manufacturer and could not be evaluated.As such, no root cause for the product malfunction could be found.A complaint history review was conducted and no similar issues have been reported against this lot number.We are awaiting receipt of one subassembly lot record to finish the product history review.As it stands, no discrepancies could be found.Once that lot record has been received, a subsequent report with the full product history review will be sent.The stryker peripheral vascular medial affairs team determined that the artix device may have caused or contributed to the patient's need for a venous cutdown.The device labeling lists vascular trauma as a possible adverse event/complication.
|
| |
|