• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMS 700 CXR MS Pump; prosthesis, penis, inflatable

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMS 700 CXR MS Pump; prosthesis, penis, inflatable Back to Search Results
Model Number 72404264
Medical Device Problem Code Human-Device Interface Problem (2949)
Health Effect - Clinical Codes Capsular Contracture (1761); Pain (1994)
Date of Event 05/21/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Model number/catalog number: 72404264.Serial number: n/a.Batch/lot number: 1100826628.Model/catalog description: cxr preconnect ms 18cm ps iz.Unique identifier (udi) #: (b)(4).
 
Event or Problem Description
It was reported that the patient implanted with an inflatable penile prosthesis (ipp) experienced scrotal pain when operating the momentary squeeze (ms) pump.A surgical procedure was performed, during which no device issue was identified.Although the pump continued to inflate the ipp as intended, its use caused significant pain to the patient.The physician noted difficulty when manipulating the ms pump and elected to replace it with a tenacio pump.The new pump was implanted in a different scrotal pocket.In addition, a capsule believed to be contributing to the patient's pain was excised.No additional patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS 700 CXR MS Pump
Common Device Name
prosthesis, penis, inflatable
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key50087091
Report Number2124215-2026-40976
Device Sequence Number10006096
Product Code JCW
UDI-Device Identifier00878953003726
UDI-Public(01)00878953003726(17)271106(10)1100826628
Combination Product (Y/N)Y
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
N970012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/26/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number72404264
Device Catalogue Number72404264
Device Lot Number1100826628
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/02/2026
Supplement Date Received by Manufacturer07/27/2026
Initial Report FDA Received Date07/30/2026
Supplement Report FDA Received Date08/26/2026
Date Device Manufactured11/06/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age60 YR
Patient SexMale
-
-