It was reported that prior to the attempted implant of a transcatheter aortic valve, a pre-implant balloon aortic valvuloplasty (bav) was performed due to the patient's calcified anatomy.Following insertion of the delivery catheter system (dcs) into the patient, on the first deployment attempt, it was observed that the valve was positioned too shallow in relation to the target implant depth.Subsequently, the valve was recaptured into the dcs.Upon the second deployment attempt, the valve was positioned too deep relative to the target implant depth.Subsequently, a second recapture was performed.On the third deployment attempt, the valve was implanted successfully, however paravalvular leak (pvl) of unspecified severity was observed.Subsequently, a post-implant bav was performed with a non-medtronic 22 millimeter (mm) balloon.The final angiogram revealed no pvl and patent flow in both coronary arteries.An echocardiogram assessment was being performed when the patient was observed to be hypotensive.Echocardiogram and coronary angiography revealed a partially obstructed left main coronary artery due, resulting in diminished flow.A stent was placed into the left main coronary artery.Per the physician, the patient's heavily calcified left main coronary artery leaflet contributed to the obstruction.
|
|
Continuation of d10: product id {d-evolutfx-2329}; product lot/serial number {unknown}; product type: {0195-heartvalves}; implant date {non-implantable} this report was impacted by emdr scheduled maintenance occurring (b)(6), 2026.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|