| Model Number |
ONYXNG25030UX |
| Medical Device Problem Codes |
Device Dislodged or Dislocated (2923); Positioning Problem (3009)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/23/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
During a procedure an attempt was made to use one onyx frontier coronary drug eluting stent when stent dislodgement occurred during removal following failed delivery.The device was inspected with no issues.Negative prep was not performed.The lesion was pre-dilated.The lesion being treated was a mildly tortuous, mildly calcified lesion located in the ostium of the left main (lm) coronary artery and circumflex (cx) artery.The device did pass through a previously-deployed stent.There was resistance encountered when advancing the device, and excessive force was not used during delivery.The stent was being delivered to the lesion site in the proximal cx artery when resistance was encountered.The stent was pulled out, and when it was pulled back it came off the balloon and got stuck in the patient in the lm and aorta.The stent was removed using a snare.No further patient complications were reported as a result of the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|