• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Anthogyr AXIOM IMPLANT ?5.2 x 6.5; Endosseous Dental Implant

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Anthogyr AXIOM IMPLANT ?5.2 x 6.5; Endosseous Dental Implant Back to Search Results
Catalog Number OP52060
Medical Device Problem Code Loss of Osseointegration (2408)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/22/2026
Type of Reportable Event Serious Injury
Event or Problem Description
The clinician reports the implant was inserted (b)(6) 2020 in fdi 47.On (b)(6) 2026, loss of osseointegration was verified.The device was forwarded to the manufacturer.There were no patient operative or post-operative complications reported.
 
Additional Manufacturer Narrative
The batch number could be verified.Our manufacturing q-system assures, that production and process controls are in place to ensure that batches confirm to the applicable specifications before they are distributed.? the loss of an endosseous dental implant after successful osseointegration and restoration is a known inherent risk of the procedure.Implants may have to be removed in case one or more of the implant success criteria are not met.Implant success criteria according to buser et al.(1991) are: 1.Absence of persistent subjective complaints such as pain, foreign body sensation and /or dysesthesia.2.Absence of a recurrent peri-implant infection with suppuration.3.Absence of implant mobility.4.Absence of a continuous radiolucency around the implant.The manufacturer's trend analysis confirms that the reported failure rate associated with its dental implants is below the expected failure rate for this treatment as published in the scientific literature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXIOM IMPLANT ?5.2 x 6.5
Common Device Name
Endosseous Dental Implant
Manufacturer (Section D)
Anthogyr
2237, avenue andr?-lasquin
sallanches 74700
FR  74700
Manufacturer (Section G)
Anthogyr
2237, avenue andr?-lasquin
sallanches 74700
FR   74700
Manufacturer Contact
jennifer jackson
60 minuteman road
andover, MA 01810
+1(978)747-2509
MDR Report Key50087874
Report Number0008020776-2026-000833
Device Sequence Number21310631
Product Code DZE
UDI-Device Identifier03663394000239
UDI-Public03663394000239
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K101913
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 07/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/26/2024
Device Catalogue NumberOP52060
Device Lot Number19-326372
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/06/2026
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Event Location Other
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/30/2026
Initial Report FDA Received Date07/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/26/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
Patient SexMale
-
-