| Model Number |
VXSK0021 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Pericardial Effusion (3271)
|
| Date of Event |
07/09/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient experienced pericardial effusion.During a pulsed field ablation (pfa) procedure for the treatment of atrial fibrillation (af), a versacross access solution was selected for use.Ultrasound guidance was used throughout the procedure.Prior to the event, the physician reported difficulty positioning a coronary sinus (cs) catheter due to challenging cs anatomy.Transseptal tsp) was successfully completed using a versacross rf wire with a single application of rf energy.No resistance was reported during guidewire advancement, no multiple attempts were required to position the versacross dilator on the septum, and no issues were reported with the rf wire during transseptal access.Immediately following completion of the tsp, intracardiac echocardiography (ice) imaging revealed a pericardial effusion that appeared to be increasing in size.An echocardiography specialist subsequently assessed the finding and reported the effusion to be associated with the coronary sinus.At the time the effusion was identified, a non-boston scientific catheter was positioned across the septum.No ablation had been performed.The procedure was subsequently terminated.The patient remained hemodynamically stable and was managed conservatively.The pericardial effusion resolved without the need for drainage.The patient was hospitalized beyond the standard course of care for additional monitoring.Following a few days of monitoring, the patient was discharged and is expected to make a full recovery.No device malfunctions were reported, and no physical damage to the device was observed prior to use.There were no known pre-existing patient conditions identified that may have contributed to the event.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient experienced pericardial effusion.During a pulsed field ablation (pfa) procedure for the treatment of atrial fibrillation (af), a versacross access solution was selected for use.Ultrasound guidance was used throughout the procedure.Prior to the event, the physician reported difficulty positioning a coronary sinus (cs) catheter due to challenging cs anatomy.Transseptal tsp) was successfully completed using a versacross rf wire with a single application of rf energy.No resistance was reported during guidewire advancement, no multiple attempts were required to position the versacross dilator on the septum, and no issues were reported with the rf wire during transseptal access.Immediately following completion of the tsp, intracardiac echocardiography (ice) imaging revealed a pericardial effusion that appeared to be increasing in size.An echocardiography specialist subsequently assessed the finding and reported the effusion to be associated with the coronary sinus.At the time the effusion was identified, a non-boston scientific catheter was positioned across the septum.No ablation had been performed.The procedure was subsequently terminated.The patient remained hemodynamically stable and was managed conservatively.The pericardial effusion resolved without the need for drainage.The patient was hospitalized beyond the standard course of care for additional monitoring.Following a few days of monitoring, the patient was discharged and is expected to make a full recovery.No device malfunctions were reported, and no physical damage to the device was observed prior to use.There were no known pre-existing patient conditions identified that may have contributed to the event.
|
| |
|
Additional Manufacturer Narrative
|
|
D4: unique identifier (udi): detailed product information was not provided to bsc.We completed a good faith effort to obtain the information, but it was not available because device was discarded.Because the product is unknown, we are unable to provide the unique identifier (udi) and other specific product information.Investigation summary:
with all the available information, boston scientific is unable to confirm cause of the reported clinical observation of a pericardial effusion.The device has not been received for analysis; therefore, a technical analysis of the complaint device could not be completed.Device history record review
no dhr review conducted as the provided upn is not available to trace back to the customer sales.Risk review:
a risk review performed for the versacross device confirmed that the event of pericardial effusion is a known event defined in the products risk management documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Labeling review:
based on the information provided, there is no evidence that the device was used in a manner inconsistent with the labelled indications/instructions for use (ifu).Investigation conclusion:
boston scientific has assigned an investigation conclusion code of no problem detected, known inherent risk of device and supporting evidence that came to this conclusion.There are several factors that may have contributed to the adverse event(s), such as excessive force during advancement of the device, physician/scrub tech technique, tortuous or challenging patient anatomy, or prior patient condition.Boston scientifics investigation findings were unable to establish a clear conclusion about the cause of the reported event.The device met all requirements prior to being approved for final distribution/sale and there was no evidence to suggest that the instructions for use (ifu) was not followed.
|
| |
|
Additional Manufacturer Narrative
|
|
D4: unique identifier (udi): detailed product information was not provided to bsc.We completed a good faith effort to obtain the information, but it was not available because device was discarded.Because the product is unknown, we are unable to provide the unique identifier (udi) and other specific product information.
|
| |
|