Catalog Number 021.3318
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Fibrosis (3167)
Date of Event 06/22/2026
Type of Reportable Event Serious Injury
Event or Problem Description
The clinician reports the implant was inserted (b)(6) 2023 in ada 25.Details of surgery: immediate extraction site.On (b)(6) 2026, fracture of the implant was verified.The device was forwarded to the manufacturer.At the event the patient experienced: mobility.No further patient complications were reported.
 
Additional Manufacturer Narrative
The batch number could be verified.Our manufacturing q-system assures, that production and process controls are in place to ensure that batches confirm to the applicable specifications before they are distributed.? the fracture of a dental implant is a known long-term complication of endosseous dental implants.The manufacturers trend analysis confirms that the reported implant fracture rate associated with its dental implants is low and remains consistent with the experience as documented in the literature.There are different reasons why implants break and this problem has been illuminated by various authors (e.G.Maeglin 1988, beumer 1989, tetsch 1991, worthington 1992).Beumer and lewis 1989 report the following causes: a) production / design flaws b) inadequate fitting of the suprastructure c) occlusal overload d) bruxism e) bone resorption around the implants f) insufficient axle alignment of the implants g) trauma h) biomechanical causes i) design of the suprastructure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLT ?3.3 NC, SLActive? 18mm, TiZr, Loxim
Common Device Name
Endosseous Dental Implant
Manufacturer (Section D)
Institut Straumann AG
peter merian-weg 12
basel 04002
SZ  04002
Manufacturer (Section G)
Institut Straumann AG
peter merian-weg 12
basel 04002
SZ   04002
Manufacturer Contact
jennifer jackson
60 minuteman road
andover, MA 01810
+1(978)747-2509
MDR Report Key50088202
Report Number0009613348-2026-011740
Device Sequence Number10479853
Product Code DZE
UDI-Device Identifier07630031729842
UDI-Public07630031729842
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K153758
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 07/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number021.3318
Device Lot NumberJTL32
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/06/2026
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Event Location Other
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/30/2026
Initial Report FDA Received Date07/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/31/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
Patient SexMale
-
-